# Known Issues
* DC offset issue. When porgram default settings, the DC offset is not the same for all channels.
* for PHA firmware, when the trigger average changed, the Events per Agg need to be changed.
* for PHA firmware, when the trigger rate changed, the Events per Agg need to be changed.
* The Agg Organization, Event per Agg, Record Length are strongly correlated. Some settings are invalid and will cause the digitizer goes crazy.
* Sometimes, the buffer is not in time order, and make the trigger/Accept rate to be nagative. This is nothing to do with the program but the digitizer settings. Recommand reporgram the digitizer.
* for 1740 QDC, RecordLenght is board setting, but readout is indivuial group.
# Known Bugs
* There is no known bug. Please report if you find one.