HV supply
The Iseg HV modules are using Mpod crate. A Mpod-C (controller) is used to control the HV module.
The Mpod-C can be controlled via ethernet, USB, or CAN. For the ethernet, it requires an IP.
GUI control and monitor
https://fsunuc.physics.fsu.edu/git/rtang/IsegSNMPGUI
The GUI also push monitor values to database and display on Grafana https://fsunuc.physics.fsu.edu/grafana/d/JG7ADC67z/iseg-hv-monitor?orgId=1&refresh=5s
IP
128.186.111.101
Iseg HV with Wiener Mpod
to control the Mpod-C via USB (say, setting up the IP), in Windows, install the MUSEController.
SNMP
More detail can be found in here.
In linux, install snmp
sudo apt install snmp
copy the WIENER-CRATE-MIB.txt into /usr/share/snmp/mibs/
Name | Type | Unit | Read/Write | Comment |
---|---|---|---|---|
outputVoltage | Float | V | R/W | Set voltage |
outputCurrent | Float | A | R/W | Set trip current |
outputMeasurementSenseVoltage | Float | V | R | Sense output voltage |
outputMeasurementCurrent | Float | A | R | Leakage current |
outputSwitch | Integer | R/W | 0 = off, 1 = on, 2 = reset Emergency Off, 3 = set emergenct Off, 10 = clear events | |
outputVoltageRiseRate | Float | V/s | R/W | |
outputVoltageFallRate | Float | V/s | R/W |
example syntax
~>snmpget -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputMeasurementSenseVoltage.u1 WIENER-CRATE-MIB::outputMeasurementSenseVoltage.u1 = Opaque: Float: 0.000000000000 V
~>snmpset -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputVoltage.u0 F 0.0 WIENER-CRATE-MIB::outputVoltage.u0 = Opaque: Float: 0.000000000000 V
Python GUI for control and monitor
https://fsunuc.physics.fsu.edu/git/rtang/IsegSNMPGUI
This program should be suitable for all Iseg HV modules using Mpod-C with SNMP protocal.
The only hard-coded thing is the Mpod-C IP, and the database IP and name.