Triton Injector: Difference between revisions

From FSU Fox's Lab Wiki
Jump to navigation Jump to search
(Created page with "The Triton Injector has components located at three different potentials with different systems: '''Ground potential:''' Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition Preaccelerator HV(-55 kV) supply Fumehood Activity Monitor Source Cooling Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehoo...")
 
mNo edit summary
 
(4 intermediate revisions by 3 users not shown)
Line 1: Line 1:
The Triton Injector has components located at three different potentials with different systems:  
The Triton Injector has components located at three different potentials with different systems:  
   
   
'''Ground potential:'''
=== Ground potential ===
Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump
 
Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition
* Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump
Preaccelerator HV(-55 kV) supply   
* Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition
Fumehood Activity Monitor
* Preaccelerator HV(-55 kV) supply   
Source Cooling
* Fumehood Activity Monitor
* Source Cooling
 
Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity,  
Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity,  
AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)   
AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)   


'''-55 kV potential:'''
=== -55 kV potential ===
Vacuum pumps: TSP 1, IGP 1  
 
Vacuum instrumentation: Ion-gauge, TC,  creating "Good vacuum 1" condition
*Vacuum pumps: TSP 1, IGP 1  
Gate-valve controller, interlocked with GND-level controller
*Vacuum instrumentation: Ion-gauge, TC,  creating "Good vacuum 1" condition
*Gate-valve controller, interlocked with GND-level controller
 
=== -65 kV potential ===
 
*Ionizer Power supply
*Extractor V Supply
*Einzel Lens Supply
*Cathode V Supply
*Immersion-Lens V Supply
*CS Boiler Supply (Variac)


'''-65 kV potential:'''
Ionizer Power supply
Extractor V Supply
Einzel Lens Supply
Cathode V Supply
Immersion-Lens V Supply
CS Boiler Supply (Variac)
All power is interlocked with GND-level Interlock controller:
All power is interlocked with GND-level Interlock controller:
Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down.  
Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down.  
All other Power-supplies are turned down fast by interlock.
All other Power-supplies are turned down fast by interlock.
[[File:Source Diagram.png|frame|left|alt=Source Diagram|Source Diagram drawn by Ben Asher]]

Latest revision as of 13:30, 14 January 2024

The Triton Injector has components located at three different potentials with different systems:

Ground potential

  • Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump
  • Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition
  • Preaccelerator HV(-55 kV) supply
  • Fumehood Activity Monitor
  • Source Cooling

Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity, AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)

-55 kV potential

  • Vacuum pumps: TSP 1, IGP 1
  • Vacuum instrumentation: Ion-gauge, TC, creating "Good vacuum 1" condition
  • Gate-valve controller, interlocked with GND-level controller

-65 kV potential

  • Ionizer Power supply
  • Extractor V Supply
  • Einzel Lens Supply
  • Cathode V Supply
  • Immersion-Lens V Supply
  • CS Boiler Supply (Variac)

All power is interlocked with GND-level Interlock controller: Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down. All other Power-supplies are turned down fast by interlock.

Source Diagram
Source Diagram drawn by Ben Asher