= Introduction =
= Introduction =
Clarion2
<ref> T. J. Gray ''et. al'', NIM A '''1041''', 167392 (2022) https://doi.org/10.1016/j.nima.2022.167392</ref>
is the successor of Clarion
<ref> C. J. Gross ''et. al'', NIM A '''450''', 12 (2000)  https://doi.org/10.1016/S0168-9002(00)00159-5</ref>
('''CL'''over '''A'''rray
for '''R'''adioactive '''ION''' beam). It is a 4π gamma ray detectors array with 16 High-Purity Germanium (HPGe) clover detectors. Each HPGe detector is shied with a BGO (Bismuth Germanate <nowiki>[</nowiki>Bi4Ge3O12<nowiki>]</nowiki>) shield.


Clarion2 is the successor of [[Clarion]]. It is a 4π gamma ray detectors array.
Each clover detector has 4 crystals, color-coded as Red, Blue, Green, and Black. Here is the angular map for all 16 clovers. Further details of the array can be found [https://doi.org/10.1016/j.nima.2022.167392 here] or [https://arxiv.org/abs/2208.05449 here].
 
[[File:Clarion2 angular map.png|400px|thumb|none|Clarion2 angular map. The purple dot is the center of each clover.]]
 
{{Notice|Solid work model needed}}


== HPGe Clover Detector ==
== HPGe Clover Detector ==
The detectors are produced by EURISYS Measure, now CANBERRA, owned by MIRION Technologies <nowiki>[</nowiki>[https://www.canberra.com/fr/produits/detectors/x-ray-detector-history.html 1]<nowiki>]</nowiki>. An approximate 3.3 kg clover detector consists of 4 coaxial crystals (dimension 50 mm in diameter, 80 mm in length) <nowiki>[</nowiki>[https://www.mirion.com/products/clover-detectors-four-coaxial-germanium-detectors Mirion webpage]<nowiki>]</nowiki>.
=== Typical photon-peak efficiency curve ===
[[File:CLARION2 efficiency curve.png|thumb|right|400px|CLARION2 efficiency curve by Tim Gray. The data points are from experimental measurments.]]
The absolute photo-peak efficiency at 1332 keV is ~1.8%.
The equation for the absolute efficiency curve is
<span style="font-size:120%">
<math> \epsilon(e_{\gamma} \text{[keV]}) = \epsilon_0 \exp^{\left( \left(A+Bx + Cx^2\right)^{-G} + \left(D+E y +F y^2\right)^{-G} \right)^{-1/G}}, </math>
</span>
<math> x = \ln\left(\frac{e_\gamma}{100}\right), y =  \ln\left(\frac{e_\gamma}{1000}\right) </math>
<math> \epsilon_0 = 0.0037679 </math>
<math> A = 13.6858 </math>
<math> B = 1.4055 </math>
<math> C = 0.000 </math>
<math> D = 3.8942 </math>
<math> E = -2.4073 </math>
<math> F = 0.0629 </math>
<math> G = 0.5000 </math>
=== GEANT4 simulation ===
The performance of a single clover detector was simulated using GEANT4. The source code can be downloaded from https://github.com/goluckyryan/HPGeClover.


[[File:GEAN4 simulation of 152Eu.png|600px|thumb|none|GEANT4 simulation of 152Eu]]
A GEANT4 simulation for Clarion2 can be found at https://fsunuc.physics.fsu.edu/git/pderosa888/CLARION2_GEANT4.
[[File:Clarion2 model.png|600px|thumb|none|Clarion2 GEANT4 model]]


== Trinity ==
== Trinity ==
[[File:Front view of TRINITY.png|400px|thumb|right|Front view of TRINITY, photo by Vandana.]]
Trinity is formed by GAGG ('''G'''adolinium '''A'''luminium '''G'''allium '''G'''arnet) scintillator crystals. It is located in the target chamber of the Clarion2. The GAGG crystals are arranged in 5 rings, covering 7-54 degrees. Currently only two rings are implemented, which cover 14-24 and 34-44 degrees respectively.


Trinity is GAGG (stand for ?) detector array.
[[File:A single GAGG detector.png|400px|thumb|none|A single GAGG detector]]


= DAQ =  
== Neutron Shield ==


In 2022 April, a set of neutron shields surrounded Clarion2 for shielding slow neutrons.


= Clarion2 private network =


The Clarion2 uses its own private network. The network structure is illustrated as following:
Fox's Lab network --- clarion2(Mac, gateway) ---- 10 Gb switch
                                                      +--------- clarion2 DAQ
                                                      +--------- LN2 control, HV control
= DAQ =
The DAQ system is fully digitized with [[Pixie16 digitizer]]. The signals from the Clover and BGO are directly fed to the Pixie16 digitizers. The data processing library developed by Tim Gray can be found at https://github.com/belmakier/libpixie and https://github.com/belmakier/libClarionTrinity .
The Plx driver was modified from the Broadcom driver. The PixieSDK (legacy) was also modified.
{{Notice | Detail of operation is needed.}}


= LN2 cooling system =
= LN2 cooling system =
The liquid Nitrogen (LN2) filling system developed by Mitch Allmond (Oak Ridge) is hosted on the DAQ computer under account "ln@ln02".
The liquid Nitrogen (LN2) filling system developed by Mitch Allmond (Oak Ridge) is hosted on the LN2-control computer under the account "ln@ln02".
== ln@ln02 Directory==
== ln@ln02 Directory==
Several important directories for using the LN2 system are listed below.  
Several important directories for using the LN2 system are listed below.  
Line 29: Line 95:
;<code> fill <det-list> </code>
;<code> fill <det-list> </code>
: fill detectors <det-list> now
: fill detectors <det-list> now
;<code> fill <det-list> after <number-of-minutes> </code> (or <hours&#58;:minutes>)  
;<code> fill <det-list> after <number-of-minutes> </code> (or <hours<nowiki>:</nowiki>minutes>)  
: wait to fill detectors <det-list> by time <hours:minutes>
: wait to fill detectors <det-list> by time <hours:minutes>
;<code> cool <det-list> </code>
;<code> cool <det-list> </code>
Line 85: Line 151:


http://clarion02.physics.fsu.edu/~clarion02/index.html
http://clarion02.physics.fsu.edu/~clarion02/index.html
= List of Collaborators =
[[File:A picture of Clarion2 collaborators.png|500px|thumb|right|From left: James Christie, Ram Yadav, Mitch Allmond, Tim Gray, photo taken by Vandana]]
The list may not be completed.
{| class="wikitable"
! Name || Affiliation
|-
| Mitch Allmond|| ORNL
|-
| Tim Gray || ORNL
|-
| Toby King || ORNL
|-
| Ingo Wiedenhoever || FSU
|-
| Sam Tabor || FSU
|-
| Vandana Tripathi || FSU
|-
| Lagy Baby || FSU
|-
|Soumik Bhattacharya || FSU
|-
| Caleb Benetti || FSU
|-
| Catur Wibisono || FSU
|-
| Samuel Ajayi || FSU
|-
| Ryan Tang || FSU
|-
| Ram Yadav || South Carolina State University
|-
| James Christie || UTK
|}
= Contact =
* Mitch Allmond mailto:allmondjm@ornl.gov
* Tim Gray mailto:graytj@ornl.gov
* Vandan Tripathi mailto:vtripath@fsu.edu
= References =

Latest revision as of 11:32, 15 February 2024

Image of Clarion2, photo by Tim.

Introduction

Clarion2 [1] is the successor of Clarion [2] (CLover Array for Radioactive ION beam). It is a 4π gamma ray detectors array with 16 High-Purity Germanium (HPGe) clover detectors. Each HPGe detector is shied with a BGO (Bismuth Germanate [Bi4Ge3O12]) shield.

Each clover detector has 4 crystals, color-coded as Red, Blue, Green, and Black. Here is the angular map for all 16 clovers. Further details of the array can be found here or here.

Clarion2 angular map. The purple dot is the center of each clover.
Notice Notice :   Solid work model needed

HPGe Clover Detector

The detectors are produced by EURISYS Measure, now CANBERRA, owned by MIRION Technologies [1]. An approximate 3.3 kg clover detector consists of 4 coaxial crystals (dimension 50 mm in diameter, 80 mm in length) [Mirion webpage].

Typical photon-peak efficiency curve

CLARION2 efficiency curve by Tim Gray. The data points are from experimental measurments.

The absolute photo-peak efficiency at 1332 keV is ~1.8%.

The equation for the absolute efficiency curve is

GEANT4 simulation

The performance of a single clover detector was simulated using GEANT4. The source code can be downloaded from https://github.com/goluckyryan/HPGeClover.

GEANT4 simulation of 152Eu

A GEANT4 simulation for Clarion2 can be found at https://fsunuc.physics.fsu.edu/git/pderosa888/CLARION2_GEANT4.

Clarion2 GEANT4 model

Trinity

Front view of TRINITY, photo by Vandana.

Trinity is formed by GAGG (Gadolinium Aluminium Gallium Garnet) scintillator crystals. It is located in the target chamber of the Clarion2. The GAGG crystals are arranged in 5 rings, covering 7-54 degrees. Currently only two rings are implemented, which cover 14-24 and 34-44 degrees respectively.

A single GAGG detector

Neutron Shield

In 2022 April, a set of neutron shields surrounded Clarion2 for shielding slow neutrons.

Clarion2 private network

The Clarion2 uses its own private network. The network structure is illustrated as following:

Fox's Lab network --- clarion2(Mac, gateway) ---- 10 Gb switch
                                                      +--------- clarion2 DAQ
                                                      +--------- LN2 control, HV control

DAQ

The DAQ system is fully digitized with Pixie16 digitizer. The signals from the Clover and BGO are directly fed to the Pixie16 digitizers. The data processing library developed by Tim Gray can be found at https://github.com/belmakier/libpixie and https://github.com/belmakier/libClarionTrinity .

The Plx driver was modified from the Broadcom driver. The PixieSDK (legacy) was also modified.

Notice Notice :   Detail of operation is needed.

LN2 cooling system

The liquid Nitrogen (LN2) filling system developed by Mitch Allmond (Oak Ridge) is hosted on the LN2-control computer under the account "ln@ln02".

ln@ln02 Directory

Several important directories for using the LN2 system are listed below.

/gln

/gln is the most important directory, it is where the control program is located and for most problems is where you will need to go

gln_ui

gln_ui is the user interface to issue commands to the fill program

gln_ui Commands

fill <det-list>
fill detectors <det-list> now
fill <det-list> after <number-of-minutes> (or <hours:minutes>)
wait to fill detectors <det-list> by time <hours:minutes>
cool <det-list>
cool down detectors
ton <det-list>
turn on (enable fill for) detectors
tof <det-list>
turn off (disable fill for) detectors
con det#[.temp#]
turn on checking of detector temps
cof det#[.temp#]
turn off checking of detector temps
abort
abort and disable all fills
load [fig-file-name]
load configuration from fig-file-name (default = gln.fig)
exit-master
exit gln_master process
Warning Warning :   This will stop the fill system
clr
clear all errors for entire system
Notice Notice :   Proper usage is to clear individual detectors
clrd <det-list>
clear errors for specific detectors
clrs
clear all slot errors
clrm
clear all manifold errors
d
display status
dt
display times of all detectors & manifolds (formatted dates and times)
dts
display times of all detectors & manifolds (in seconds)
dc
display configuration (valve and ADC IDs, etc)
mon
enter monitor mode; type control-C to exit
quit
exit gln_ui process
hvln
useless command for backward compatibility
clear
clear screen
 ? or help
help text for commands

Low Level Commands

radc <#>
read ADC number <#>
opv <#> clv <#>
open and close valves
Warning Warning :   Be careful with this

http://clarion02.physics.fsu.edu/~clarion02/index.html

List of Collaborators

From left: James Christie, Ram Yadav, Mitch Allmond, Tim Gray, photo taken by Vandana

The list may not be completed.

Name Affiliation
Mitch Allmond ORNL
Tim Gray ORNL
Toby King ORNL
Ingo Wiedenhoever FSU
Sam Tabor FSU
Vandana Tripathi FSU
Lagy Baby FSU
Soumik Bhattacharya FSU
Caleb Benetti FSU
Catur Wibisono FSU
Samuel Ajayi FSU
Ryan Tang FSU
Ram Yadav South Carolina State University
James Christie UTK