= HV supply =


The Iseg HV modules are using Mpod crate. A Mpod-C (controller) is used to control the HV module.
The Mpod-C can be controlled via ethernet, USB, or CAN. For the ethernet, it requires an IP. 
== GUI control and monitor ==
[[File:IsegHVContol Screenshot.png|thumb]]
https://fsunuc.physics.fsu.edu/git/rtang/IsegSNMPGUI
The GUI also push monitor values to database and display on Grafana https://fsunuc.physics.fsu.edu/grafana/d/JG7ADC67z/iseg-hv-monitor?orgId=1&refresh=5s
== IP ==
128.186.111.101
== Iseg HV with Wiener Mpod ==
to control the Mpod-C via USB (say, setting up the IP), in Windows, install the MUSEController.
== SNMP ==
More detail can be found in [[:File:Iseg SNMP Programmers Guide.pdf | here]].
In linux, install snmp
sudo apt install snmp
copy the [[:File:WIENER-CRATE-MIB.txt|WIENER-CRATE-MIB.txt]] into /usr/share/snmp/mibs/
{|class='wikitable'
! Name !! Type !! Unit !! Read/Write  !! Comment
|-
|outputVoltage || Float || V || R/W || Set voltage
|-
|outputCurrent || Float || A || R/W || Set trip current
|-
|outputMeasurementSenseVoltage || Float || V || R || Sense output voltage
|-
|outputMeasurementCurrent || Float || A || R || Leakage current
|-
|outputSwitch || Integer || || R/W || 0 = off, 1 = on, 2 = reset Emergency Off, 3 = set emergenct Off, 10 = clear events
|-
|outputVoltageRiseRate || Float || V/s || R/W ||
|-
|outputVoltageFallRate || Float || V/s || R/W ||
|}
=== example syntax ===
~>snmpget -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputMeasurementSenseVoltage.u1
WIENER-CRATE-MIB::outputMeasurementSenseVoltage.u1 = Opaque: Float: 0.000000000000 V
~>snmpset -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputVoltage.u0 F 0.0
WIENER-CRATE-MIB::outputVoltage.u0 = Opaque: Float: 0.000000000000 V
=== Python GUI for control and monitor ===
https://fsunuc.physics.fsu.edu/git/rtang/IsegSNMPGUI
This program should be suitable for all Iseg HV modules using Mpod-C with SNMP protocal.
The only hard-coded thing is the Mpod-C IP, and the database IP and name.