  ~>snmpget -v 2c -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputMeasurementSenseVoltage.u1
  ~>snmpget -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputMeasurementSenseVoltage.u1
  WIENER-CRATE-MIB::outputMeasurementSenseVoltage.u1 = Opaque: Float: 0.000000 V
  WIENER-CRATE-MIB::outputMeasurementSenseVoltage.u1 = Opaque: Float: 0.000000000000 V


  ~>snmpset -v 2c -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputVoltage.u0 F 0.0
  ~>snmpset -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputVoltage.u0 F 0.0
  WIENER-CRATE-MIB::outputVoltage.u0 = Opaque: Float: 0.000000 V
  WIENER-CRATE-MIB::outputVoltage.u0 = Opaque: Float: 0.000000000000 V
 
More detail can be found in [[:File:Iseg SNMP Programmers Guide.pdf here]].
.

Revision as of 20:43, 14 July 2022

HV supply

The Iseg HV modules are using Mpod crate. A Mpod-C (controller) is used to control the HV module.

The Mpod-C can be controlled via ethernet, USB, or CAN. For the ethernet, it requires an IP.

GUI control and monitor

https://fsunuc.physics.fsu.edu/git/rtang/IsegSNMPGUI

The GUI also push monitor values to database and display on Grafana https://fsunuc.physics.fsu.edu/grafana/d/JG7ADC67z/iseg-hv-monitor?orgId=1&refresh=5s

IP

128.186.111.101

Iseg HV with Wiener Mpod

to control the Mpod-C via USB (say, setting up the IP), in Windows, install the MUSEController.

SNMP

in linux, install snmp

sudo apt install snmp

copy the WIENER-CRATE-MIB.txt into /usr/share/snmp/mibs/

example syntax

~>snmpget -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputMeasurementSenseVoltage.u1
WIENER-CRATE-MIB::outputMeasurementSenseVoltage.u1 = Opaque: Float: 0.000000000000 V
~>snmpset -v 2c -Op .12 -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputVoltage.u0 F 0.0
WIENER-CRATE-MIB::outputVoltage.u0 = Opaque: Float: 0.000000000000 V

More detail can be found in File:Iseg SNMP Programmers Guide.pdf here. .