Triton Injector: Difference between revisions

From FSU Fox's Lab Wiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 2: Line 2:
   
   
'''Ground potential:'''  
'''Ground potential:'''  
Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump\\
 
Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump
 
Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition
Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition
Preaccelerator HV(-55 kV) supply   
Preaccelerator HV(-55 kV) supply   
Fumehood Activity Monitor
Fumehood Activity Monitor
Source Cooling
Source Cooling
Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity,  
Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity,  
AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)   
AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)   


'''-55 kV potential:'''  
'''-55 kV potential:'''  
Vacuum pumps: TSP 1, IGP 1  
Vacuum pumps: TSP 1, IGP 1  
Vacuum instrumentation: Ion-gauge, TC,  creating "Good vacuum 1" condition
Vacuum instrumentation: Ion-gauge, TC,  creating "Good vacuum 1" condition
Gate-valve controller, interlocked with GND-level controller
Gate-valve controller, interlocked with GND-level controller


'''-65 kV potential:'''
'''-65 kV potential:'''
Ionizer Power supply
Ionizer Power supply
Extractor V Supply
Extractor V Supply
Einzel Lens Supply
Einzel Lens Supply
Cathode V Supply
Cathode V Supply
Immersion-Lens V Supply  
Immersion-Lens V Supply  
CS Boiler Supply (Variac)  
CS Boiler Supply (Variac)  
All power is interlocked with GND-level Interlock controller:
All power is interlocked with GND-level Interlock controller:
Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down.  
Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down.  
All other Power-supplies are turned down fast by interlock.
All other Power-supplies are turned down fast by interlock.

Revision as of 11:38, 3 May 2022

The Triton Injector has components located at three different potentials with different systems:

Ground potential:

Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump

Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition

Preaccelerator HV(-55 kV) supply

Fumehood Activity Monitor

Source Cooling

Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity, AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)

-55 kV potential:

Vacuum pumps: TSP 1, IGP 1

Vacuum instrumentation: Ion-gauge, TC, creating "Good vacuum 1" condition

Gate-valve controller, interlocked with GND-level controller

-65 kV potential:

Ionizer Power supply

Extractor V Supply

Einzel Lens Supply

Cathode V Supply

Immersion-Lens V Supply

CS Boiler Supply (Variac)

All power is interlocked with GND-level Interlock controller: Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down. All other Power-supplies are turned down fast by interlock.