HV supply
IP
128.186.111.101
Iseg HV with Wiener Mpod
The Iseg HV modules are using Mpod crate. A Mpod-C (controller) is used to control the HV module.
The Mpod-C can be controlled via ethernet, USB, or CAN. For the ethernet, it requires an IP.
to control the Mpod-C via USB (say, setting up the IP), in Windows, install the MUSEController.
SNMP
in linux, install snmp
sudo apt install snmp
copy the WIENER-CRATE-MIB.txt into /usr/share/snmp/mibs/
example syntax
~>snmpget -v 2c -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputMeasurementSenseVoltage.u1 WIENER-CRATE-MIB::outputMeasurementSenseVoltage.u1 = Opaque: Float: 0.000000 V
~>snmpset -v 2c -m +WIENER-CRATE-MIB -c guru 128.186.111.101 outputVoltage.u0 F 0.0 WIENER-CRATE-MIB::outputVoltage.u0 = Opaque: Float: 0.000000 V