The Triton Injector has components located at three different potentials with different systems:

Ground potential

  • Vacuum pumps: Titanium-Sublimation Pump (TSP) 2, Ion-getter pump (IGP)2, Turbo Pump, Roughing pump
  • Vacuum instrumentation: Ion-gauge, 2 TC, creating "Good vacuum 2" condition
  • Preaccelerator HV(-55 kV) supply
  • Fumehood Activity Monitor
  • Source Cooling

Interlock controller (monitoring Vacuum status from GND, -55 kV, monitoring cooling status, fumehood activity, AC power status) (providing shutdown signal for source Gate-valve, ion source power, Preaccelerator HV.)

-55 kV potential

  • Vacuum pumps: TSP 1, IGP 1
  • Vacuum instrumentation: Ion-gauge, TC, creating "Good vacuum 1" condition
  • Gate-valve controller, interlocked with GND-level controller

-65 kV potential

  • Ionizer Power supply
  • Extractor V Supply
  • Einzel Lens Supply
  • Cathode V Supply
  • Immersion-Lens V Supply
  • CS Boiler Supply (Variac)

All power is interlocked with GND-level Interlock controller: Ionizer Power Supply has Battery-backup and "10 second" interlock turn-down. All other Power-supplies are turned down fast by interlock.

Source Diagram
Source Diagram drawn by Ben Asher