Split-Pole Spectrograph: Difference between revisions

From FSU Fox's Lab Wiki
Jump to navigation Jump to search
Line 17: Line 17:


== Focal plane detector ==
== Focal plane detector ==
[[File:Focal plane detector.png|400px|thumb|right|Front view of the SPS focal plane detector with the front window removed.]]
[[File:Focal plane detector.png|400px|thumb|right|Front view of the opened camerabox. The SPS focal plane detector with the front window removed is at the bottom.]]


[[File:PID EDE annoteted.png|400px|thumb]]
[[File:PID EDE annoteted.png|400px|thumb]]

Revision as of 23:17, 27 May 2023

Annotated picture of the SE-SPS, An plain picture is here : File:SPS Magnet.png
SE-SPS COSY simulation. An plain picture is here File:SPS Sketch With Cosy.png

The Super Enge Split-Pole Spectrograph [1] [2] [3] was originally located at the Wright Nuclear Structure Laboratory (closed at 2013), at Yale University. It was moved to FSU in the fall of 2013. It consists of a reaction chamber, a split-pole magnetic spectrograph, a position-sensitive ionization drift chamber, and a plastic scintillator. It has an angular acceptance of 128 msr (vertical ±40 mrad, horizontal ±80 mrad). The maximum B-field is 1.63 T with a radius of curvature from 511 mm to 920 mm. The mean radius is 600 mm.

The Super Enge Split-Pole Spectrograph is an upgrade of the Yale Enge SPS. The major change is the redesign of the backward silicon detector array to the SABRE.

Magnet

The SPS contains 2 dipole magnets: pole-1 and pole-2. The magnet can be rotated from 0 to 55 degrees in the lab. The magnetic field has an upper limit of 1.63 T (or 16.3 kG).

Focal plane detector

Front view of the opened camerabox. The SPS focal plane detector with the front window removed is at the bottom.
PID EDE annoteted.png

The focal plane detector [4] [5] consists of an ion drift chamber with a set of delay lines to detect the position of a particle along the focal plane and a plastic scintillator to detect the energy of the incoming particle. Using the energy loss of the particle through the ion chamber with the energy deposited in the scintillator, particles of different charges and masses can be identified.

The typical pressure of the drift chamber is 70 to 300 Torr of isobutane gas [HC(CH3)3]. The pressure controls the density of the gas and affects the bias voltage, it further affects the drift velocity. Here is a table of pressure and bias voltages (data taken from the Ph.D. thesis of Erin Good).

Gas pressure (Torr) Anode bias (V) Cathode plate bias (V)
70 +1050 to +1035 -550 to -500
80 +1150 -550
100 +1250 -600
110 +1200 to 1320 -620 to -600
125 +1425 -650
130 +1360 -725
150 +1500 -700

From bottom to top, the cathode plate, drift region (contains four biased field-shaping wire grids), Frisch grid (grounded), three anode wires, and pickup pads (which are with the delay lines). Electrons induced by any radiation will drift upward, pass the Frisch grid, are accelerated by the anodes, and hit the pickup pads. The pickup pads are strips with 45° against the anode wires, almost parallel to the particle trajectories[5]. Each pickup strip is 0.09" (2.286 mm) wide and 1.4" (35.56 mm) long, and spaced 0.01" (0.245 mm). A total of 440 strips with a 5 ns delay per strip results in a nominal total delay of 2.2 μs. The position of the hit position can then be determined by the time difference at the end of the delay line.

There are two position-sensitive delay lines in the focal plane detector. By reconstructing the particle trajectory using the position information of both delay lines, the resolution can be enhanced by correcting for the kinematic shift of the reaction.

AnnotatedFocalPlaneRay.png
FPShift.gif
Simulated rays near the focal plane. An animation on the shift of the focal panel. An optimum is reached at FP = -42 mm.

After passing the drift chamber, the particles will be stopped and detected in a plastic scintillator with a photomultiplier tube (PMT) at each end. Together with the energy loss, obtained by the cathode in the drift chamber, a ΔE-E particle identification can be done.


SABRE

Installing particle shield on SABER (photo taken on May 5, 2022)

The Silicon Array for Branching Ratio Experiments

SABRE is a silicon array designed around branching ratio experiments with the SPS. SABRE sits at backwards angles from the target, and covers roughly 30% of 4pi. SABRE has both thick and thin dead-layer detectors, with the thin dead-layer detectors capable of reaching ~200 keV thresholds for protons and deuterons.

https://www.sciencedirect.com/science/article/pii/S0168900221002837

SPS Experiment Guide

Media:SPS_Experiment_Guide.pdf

SPS Operating Procedures

I created this section as a place to store procedures for the chamber swaps, however, I expect there are other things we might want to document here. -p

Repositories

https://github.com/sesps

Contact

References

  1. H.A. Enge, NIM 162, 161 (1979) https://doi.org/10.1016/0029-554X(79)90711-0
  2. H. A. Enge, NIM 187, 1 (1981) https://doi.org/10.1016/0029-554X(81)90465-1
  3. J. E. Spencer and H. A. Enge, NIM 49, 181 (1967) https://doi.org/10.1016/0029-554X(67)90684-2
  4. C. Marshal et. al, IEEE Tran. Inst. and Meas. 68, 533 (2018) https://doi.org/10.1109/TIM.2018.2847938
  5. 5.0 5.1 R. G. Markham and R. G. H. Robertson, NIM 129, 131 (1975) https://doi.org/10.1016/0029-554X(75)90122-6