Split-Pole Spectrograph
Notice : | Need a picture of SPS |
The Super Enge Split-Pole Spectrograph [1] [2] is a magnet spectrometer to measure the spectrum of nuclear reactions. The concept and design were developed by Harald A. Enge[3] at 1967, aimed to have a broad-momentum range spectrograph with or . The spectrometer was originally located at the Wright Nuclear Structure Laboratory (closed at 2013), at Yale University. It was moved to FSU in the fall of 2013. It consists of a reaction chamber, a split-pole magnetic spectrograph, a position-sensitive ionization drift chamber, and a plastic scintillator. It has an angular acceptance of 128 msr (vertical ±40 mrad, horizontal ±80 mrad). The maximum B-field is 1.63 T with a radius of curvature from 511 mm to 920 mm. The mean radius is 600 mm. The advantage of the split-pole instead of a single-pole magnet is the aberration (x|θ2) and (x|φ2) are almost zero [1].
The Super Enge Split-Pole Spectrograph is an upgrade of the Yale Enge SPS. The major change is the redesign of the backward silicon detector array to the SABRE.
Magnet
A general discussion of magnetic spectrographs can be found at Ref. [1] and Ref. [4]. There are many designs from a simple single dipole to a combination of multiple dipoles and quadrupoles.
The SPS magnet was designed for a large solid angle, large resolving power, and correction of kinematic broadening. Using two-directional focusing and second-order focusing spectrograph can achieve a large sold angle and resolving power. Second-order focusing means the second-order terms in the acceptance angles vanish, i.e. no aberration.
The SPS contains 2 separate poles enveloped by a single coil. The split provides second-order double focusing over a broad range of momenta. The magnet can be rotated from 0 to 55 degrees in the lab. The magnetic field has an upper limit of 1.63 T (or 16.3 kG).
Property | Symbol | Value |
---|---|---|
Orbital radius | 511 to 920 mm | |
Resolving power | 1st order of 4290 (at mrad) | |
Acceptance | Horizontal | 160 mrad |
Vertical | 80 mrad | |
Dispersion | 1.96 | |
Magnification | 0.39 | |
2.9 | ||
Maximum field | 1.63 T |
Transfer matrix and COSY INFINITY simulation
Notice : | need to fill up. Any 1st few orders transfer matrixes? |
The entrance coordinates of the beam are wiht momentum , and coordinate at exit are . The entrance and exit coordinates are related by
using Taylor expansion:
In the above expansion, the term is the magnification in the x-direction. is the dispersion, and is aberration. The focal plane is the z-position that , i.e. the exit does not depend on the entrance angle.
Kinematic broadening
Kinematic broadening is the broadening of focus for the same reaction state that the term is not zero. After a reaction, the angle and momentum of the recoil particle are related that the entrance angle is a function of momentum. For each energy state, the relation between the angle and momentum is unique. For example, in a 2-body transfer reaction, the momentum vector is
where are the momentum and the scattering angle at the CM frame, are the Lorentz factor from Lab frame to CM frame, and is the mass of the particle. All 5 coefficients are constant for a fixed energy state. And the different state is characterized by . Defined the kinematic factor K:
The kinematic broadening can be corrected by shifting the focal plane by
Focal plane detector
Notice : | The drift ion chamber was repaired in summer 2018 |
The focal plane detector [5] [6] consists of an ion drift chamber with a set of delay lines to detect the position of a particle along the focal plane and a plastic scintillator to detect the energy of the incoming particle. Using the energy loss of the particle through the ion chamber with the energy deposited in the scintillator, particles of different charges and masses can be identified.
The typical pressure of the drift chamber is 70 to 300 Torr of isobutane gas [HC(CH3)3]. The pressure controls the density of the gas and affects the bias voltage, it further affects the drift velocity.
Gas pressure (Torr) | Anode bias (V) | Cathode plate bias (V) |
---|---|---|
70 | +1050 to +1035 | -550 to -500 |
80 | +1150 | -550 |
100 | +1250 | -600 |
110 | +1200 to 1320 | -620 to -600 |
125 | +1425 | -650 |
130 | +1360 | -725 |
150 | +1500 | -700 |
From bottom to top, the cathode plate, drift region (contains four biased field-shaping wire grids), Frisch grid (grounded), three anode wires, and pickup pads (which are with the delay lines). Electrons induced by any radiation will drift upward, pass the Frisch grid, are accelerated by the anodes, and hit the pickup pads. The pickup pads are strips with 45° against the anode wires, almost parallel to the particle trajectories[6]. Each pickup strip is 0.09" (2.286 mm) wide and 1.4" (35.56 mm) long, and spaced 0.01" (0.245 mm). A total of 440 lead-coated copper strips with a 5 ns delay per strip results in a nominal total delay of 2.2 μs. Every 10 strips share a delay chip. The position of the hit position can then be determined by the time difference at the end of the delay line.
There are two position-sensitive delay lines (separated by 42.8625 mm) in the focal plane detector. By reconstructing the particle trajectory using the position information of both delay lines, the resolution can be enhanced by correcting for the kinematic shift of the reaction.
After passing the drift chamber, the particles will be stopped and detected in a plastic scintillator with a photomultiplier tube (PMT) at each end. Together with the energy loss, obtained by the cathode in the drift chamber, a ΔE-E particle identification can be done.
Outline of the algorithm
There are 7 readouts channels from the focal plane detector:
readout | type of signal |
---|---|
cathode | energy loss |
Front delay line Left | timing |
Front delay line Right | timing |
Rear delay line Left | timing |
Rear delay line Right | timing |
PMT Left | energy |
PMT Right | energy |
Kinematic correction of the focal plane
As pointed out before, the kinematic broadening can be corrected. In SPS, the dispersion D is 1.96, magnification is 0.39.
Simulated rays near the focal plane. | An animation on the shift of the focal panel. An optimum is reached at FP = -42 mm. |
A parallel shift of the focal plane maybe not be enough. Suppose the best focal plan is given by a function . The 2 positions extracted from the front and rear delay lines are , and the distance between the front and rear delay lines is . The X-avg is the solution of the equation:
For a linear tilted plane , the X-avg is
SABRE
SABRE is a Silicon Array for Branching Ratio Experiments [7] with the SPS. Its predecessor is the Yale Lamp Shade Array (YLSA). SABRE sits at backward angles from the target and covers roughly 30% of 4π. SABRE has both thick and thin dead-layer detectors, with the thin dead-layer detectors capable of reaching ~200 keV thresholds for protons and deuterons.
SPS Experiment Guide
Media:SPS_Experiment_Guide.pdf
SPS Operating Procedures
I created this section as a place to store procedures for the chamber swaps, however, I expect there are other things we might want to document here. -p
Repositories
Contact
- Jeff Blackmon mailto:blackmon@lsu.edu
- Ingo
- who should be contacted?
References
- ↑ 1.0 1.1 1.2 1.3 H.A. Enge, NIM 162, 161 (1979) https://doi.org/10.1016/0029-554X(79)90711-0
- ↑ H. A. Enge, NIM 187, 1 (1981) https://doi.org/10.1016/0029-554X(81)90465-1
- ↑ J. E. Spencer and H. A. Enge, NIM 49, 181 (1967) https://doi.org/10.1016/0029-554X(67)90684-2
- ↑ H. A. Enge, Physics Today 20, 65 (1967) https://doi.org/10.1063/1.3034401
- ↑ C. Marshal et. al, IEEE Tran. Inst. and Meas. 68, 533 (2018) https://doi.org/10.1109/TIM.2018.2847938
- ↑ 6.0 6.1 R. G. Markham and R. G. H. Robertson, NIM 129, 131 (1975) https://doi.org/10.1016/0029-554X(75)90122-6
- ↑ E. C. Good et. al, NIM A 1003, 165299 (2021) https://www.sciencedirect.com/science/article/pii/S0168900221002837